tsufz commentedSep 25, 2018
Hi,
As follow up to a discussion of Herbert and Martin, they suggest to include the scan range as a new tag in the record format, e.g. AC$MASS_SPECTROMETRY: SCAN_RANGE MS2 100-1000
Alternatively, it could be the smallest scan mass used such as:
AC$MASS_SPECTROMETRY: MIN_SCAN_Mass MS2 100
This is especially important for Orbitrap CID spectra, in which small ions maybe discriminated because of the lower mass cutoff (1/3 precursor mass). Those peaks are available in similar spectra of QTofs oder QExs. The consequence is a lower match score. In the follow up it would be make sense to adapt the search filter in order to allow filtering to the real scan window. An example, I acquire a scan range 50-100 m/z with peaks at 75-80 m/z, but search only in the range 100-300 m/z which may get a better match.